●Compact, integrated design for easy installation, reduced size, and lower transportation costs.
●User-friendly touchscreen operation requiring minimal training; daily maintenance completes in 5 minutes.
●Independent wastewater/liquid separation channels with verification, supporting dynamic control technical requirements.
●Built-in 220V AC pump control interface (no external relay required).
●Door opening time and count logging functionality.
●Leak detection, optional reagent management, and reagent refrigeration capabilities.
●Automatic alarm output for abnormal or exceeding standard measurement results.
●Timed cleaning and measurement functions; automatic post-measurement cleaning of processing units, pipelines, valves, etc.
●Multiple signal outputs: 4-20 mA, RS485/RJ45.
●Manual, timed, and automatic calibration modes, including external trigger calibration; configurable calibration intervals.
●Self-diagnostic functions, power failure protection, and automatic recovery upon power restoration.
●Robust data processing and storage, retaining raw data and operation logs for over 1 year.
Parameter |
Technical Specifications |
Detection Principle |
Zinc Reagent Spectrophotometric Method |
Application Scope |
Surface water, groundwater, industrial wastewater, domestic sewage |
Measurement Range |
0–2/10/20 mg/L (extendable) |
Indicated Error |
±5% |
Quantitation Limit |
≤0.03 mg/L |
Precision |
≤5% |
Zero Drift |
±5% |
Span Drift |
±5% |
Spike Recovery Test |
90%–110% |
Real Sample Comparison |
≤0.1 mg/L (zinc <0.2 mg/L); ≤10% (zinc ≥0.2 mg/L) |
Minimum Maintenance Cycle |
≥720 h/occurrence |
Measurement Time |
50 minutes |