●Compact, integrated design for easy installation, reduced size, and lower transportation costs.
●User-friendly touchscreen operation requiring minimal training; daily maintenance completes in 5 minutes.
●Independent wastewater/liquid separation channels with verification, supporting dynamic control technical requirements.
●Built-in 220V AC pump control interface (no external relay required).
●Door opening time and count logging functionality.
●Leak detection, optional reagent management, and reagent refrigeration capabilities.
●Automatic alarm output for abnormal or exceeding standard measurement results.
●Timed cleaning and measurement functions; automatic post-measurement cleaning of processing units, pipelines, valves, etc.
●Multiple signal outputs: 4-20 mA, RS485/RJ45.
●Manual, timed, and automatic calibration modes, including external trigger calibration; configurable calibration intervals.
●Self-diagnostic functions, power failure protection, and automatic recovery upon power restoration.
●Robust data processing and storage, retaining raw data and operation logs for over 1 year.
Parameter |
Technical Specifications |
Application Scope |
Online automatic monitoring of total iron in surface water, groundwater, industrial wastewater, and domestic sewage |
Measurement Principle |
1,10-Phenanthroline Spectrophotometric Method |
Measurement Range |
0–5 mg/L; 0–25 mg/L; 0–100 mg/L; 0–500 mg/L (extendable or customizable) |
Precision |
≤5% |
Detection Limit |
0.03 mg/L |
Zero Drift |
±5% F.S |
Span Drift |
±5% F.S |
Real Sample Comparison |
≤15% |
MTBF |
≥720 h/occurrence |
Analysis Time |
≤30 minutes |
Minimum Maintenance Cycle |
Reagent replacement every month |
Auto-Calibration |
Online auto-calibration and verification |
Dimensions |
380 × 310 × 550 mm (W × D × H) |
Weight |
16 kg |
Maintenance |
Tubing and valve replacement in 5 minutes; one-click operation post-maintenance; online software upgrades; detailed operation logs for traceability |
Data Storage |
500,000 historical data entries (lifetime full-data recording) |
Environmental Conditions |
Temperature: 5–45 ℃; Humidity ≤90% (non-condensing) |
Power Supply |
220V AC ±10%, 50–60 Hz |