Zn Total Zinc Online Analyzer

Product Description
Application
Product Features
Specification Parameters
Download

【Product Description】

The Zn Total Zinc Water Quality Auto Analyzer employs the zinc reagent spectrophotometric method to automate the measurement of total zinc in water samples. Under high-temperature and high-pressure conditions, zinc is separated from the sample. In an alkaline medium, zinc ions react with the zinc reagent to form a blue complex, which exhibits stable maximum absorbance at 610 nm. Within a specific concentration range, the absorbance is proportional to the zinc concentration, forming the quantitative basis for online zinc monitoring.

【Application】

Widely used for monitoring zinc in industrial wastewater, domestic sewage, and surface water.

【Product Features】

Compact, integrated design for easy installation, reduced size, and lower transportation costs.

User-friendly touchscreen operation requiring minimal training; daily maintenance completes in 5 minutes.

Independent wastewater/liquid separation channels with verification, supporting dynamic control technical requirements.

Built-in 220V AC pump control interface (no external relay required).

Door opening time and count logging functionality.

Leak detection, optional reagent management, and reagent refrigeration capabilities.

Automatic alarm output for abnormal or exceeding standard measurement results.

Timed cleaning and measurement functions; automatic post-measurement cleaning of processing units, pipelines, valves, etc.

Multiple signal outputs: 4-20 mA, RS485/RJ45.

Manual, timed, and automatic calibration modes, including external trigger calibration; configurable calibration intervals.

Self-diagnostic functions, power failure protection, and automatic recovery upon power restoration.

Robust data processing and storage, retaining raw data and operation logs for over 1 year.

【Specification Parameters】

Parameter

Technical Specifications

Detection Principle

Zinc Reagent Spectrophotometric Method

Application Scope

Surface water, groundwater, industrial wastewater, domestic sewage

Measurement Range

0–2/10/20 mg/L (extendable)

Indicated Error

±5%

Quantitation Limit

≤0.03 mg/L

Precision

≤5%

Zero Drift

±5%

Span Drift

±5%

Spike Recovery Test

90%–110%

Real Sample Comparison

≤0.1 mg/L (zinc <0.2 mg/L); ≤10% (zinc ≥0.2 mg/L)

Minimum Maintenance Cycle

≥720 h/occurrence

Measurement Time

50 minutes


Leave a message

By supplying my contact information, I authorize Linksens to contact me about relevant products and services.

Related Products

Customer Reviews

5.0
Based on 0 reviews
Write A Review
Rating
All ratings

Share your thoughts

Scoring *
Write a review *
Add a headline *
Your name *
(Extremely Likely)
Your email address *
We'll send you an email to verify this review camefrom you.

If applicable, please rate your satisfaction with our Linksens Service Team

(Customer Service, Tech Support, Applications). Choose 1

*reguired fields
×

Please leave message to get a quotation and solution

VerifyCode *
*reguired fields