●Compact, integrated design for easy installation and integration into diverse applications.
●User-friendly touchscreen operation requiring minimal training; daily maintenance completes in 5 minutes.
●Independent wastewater/liquid separation channels with verification, supporting dynamic control technical requirements.
●Built-in 220V AC pump control interface (no external relay required).
●Door opening time and count logging functionality.
●Leak detection, optional reagent management, and reagent refrigeration capabilities.
●Automatic alarm output for abnormal or (exceeding standard) measurement results.
●Timed cleaning and measurement functions; automatic post-measurement cleaning of processing units, pipelines, valves, etc.
●Multiple signal outputs: 4-20 mA, RS485/RJ45.
●Manual, timed, and automatic calibration modes, including external trigger calibration; configurable calibration intervals.
●Self-diagnostic functions, power failure protection, and automatic recovery upon power restoration.
●Robust data processing and storage, retaining raw data and operation logs for over 1 year.
Parameter |
Technical Specifications |
Measurement Method |
Diazo Coupling Spectrophotometry |
Measurement Range |
0–2/10/20/50 mg/L (customizable) |
Indicated Error |
±10% |
Zero Drift |
±5% |
Span Drift |
±5% |
MTBF |
≥720 h/occurrence |
Repeatability |
±3% |
Real Sample Comparison |
±10% |
Minimum Maintenance Cycle |
≥168 h/occurrence |
Measurement Time |
≤40 minutes |
Output |
Communication: 1× RS232, 1× RS485 (MODBUS protocol) |
Network: RJ45 |
|
Digital Outputs: 2 channels |
|
Digital Inputs: 1 channel |
|
Sampling Pump Relay: 1× (220V control output, pump power <1100W) |
|
Analog Output: 1× (4-20 mA) |
|
Dimensions |
380 × 310 × 550 mm (W × D × H) |
Weight |
16 kg |